CPL8100A (100A/600kHz)

Band width (-3dB)  600kHz

Current transfer ratio  0.1V/A




浏览次数:819
说明书: 产品说明书

1. Summary

CPL8100A/B is a kind of current probe that can measure both DC and AC current up to 100A (70.7Arms). The band width of CPL8100A is 600 kHz (-3dB), and the band width of CPL8100B is 2MHz (-3dB).CPL8100 series provide two range option of 10A and 100A, the users may choose the correct range according to the current value. The auto zero function make it easy for the users to operate, and it also has power and low voltage indicator as well as overload sound alert. The probe can be powered by battery or external power supply, which makes the measurement a lot more convenient. With standard BNC output port, the probe can connect to any oscilloscope, and it can also connect to multimeter with BNC-banana plug. The CPL8100 series is usually used in  power supply , motor drive and power supply. 


Adapter CK-612
BNC Output cable CK-310/320
9V battery

Electronic Characteristics

Test Condition: 23, 60%RH, cable under test get through the test center, load resistance 1MΩ

 

Range level

L

H

Current range

50mA~10A  Peak

1A~100A  Peak

Attenuation accuracy

0.1V/A

0.01V/A

Typical DC precision

3%±50mA

500mA~40A Peak 4%±50mA40A~100A Peak ±15% Maximum

Band-width

(-3dB)

CPL8100A

DC-600kHz Refer to the typical Amplitude frequency characteristic (Figure 1)

CPL8100B

DC-2MHz Refer to the typical Amplitude frequency characteristic (Figure 2)

Max current VS frequency

characteristics curve

Figure 3

Phase shift

DC65Hz<1.5°

DC65Hz<1°

Typical DC linearity

The typical DC linearity at H level(0.01V/A), Figure 4

Rise time

CPL8100A

≤583ns

CPL8100B

≤175ns

Max operation current

Please refer to voltage and current rating table

Max operation voltage

Please refer to voltage and current rating table

Max floating voltage

Please refer to voltage and current rating table

Operating voltage RMS

CATI 600V CATII 600V CATIII 300V

Common mode voltage RMS

CATI 600V CATII 600V CATIII 300V

Typical battery type and life.

9V alkaline layer-built battery/ 15H

Low power indication

When battery voltage is lower than 6.5V, battery indicator will turned red and alert.

Overload indication

When the current under test surpasses the range, the buzzer will buzz