Frequency range 9kHz—30MHz
AC voltage 0~264VAC
frequency range 50~60Hz ±5%
DC voltage range 0~55V DC
1. Introduction
EM5040/5040A/5040B Line Impedance Stabilization Network is a kind of (50uH+5Ω)||50Ω V type LISN. This product can provide stable impedance from 9 kHz to 30 MHz RF range for DUT and reference. At the same time, it isolated the useless signal from the electric network to measurement circuit and coupling the interference voltage of the DUT to the input port of the measurement receiver. The product’s performance is compliant with standard requirements of CISPR16-1-2 for V-networks. It suits conducted disturbance voltage measurement for single phase equipment well. The standard BNC output port and 50Ω output impedance make the device compatible to the measurement equipment such as receiver and spectrum analyzer of any manufacturer. The artificial hand of the product can simulate the handle device to measure.
1. Electronic Specifications
EM5040 | |
Frequency range | 9kHz—30MHz |
Circuit type | (50uH+5Ω)|| 50Ω Type V |
Measurement circuit phase number | Single phase two wire L/N1 |
AC voltage/ frequency range | 0~264VAC / 50~60Hz ±5% |
DC voltage range | 0~55V DC |
Power supply output port | German standard power supply socket |
Output current | Rating 16A |
Ground protection |
Side metal panel grounding Rear panel 4mm butterfly bolt grounding |
Output port | Standard 50Ω BNC female, |
Artificial hand | Yes |
Output signal filter | No |
Output signal limiter | No |
Output signal common mode and differential mode separator | No |
Operating temperature range | 0℃~45℃ |
Storing temperature | -20℃~70℃ |
Safety standard | EN61010-1 |
Size | 338mm(Length)×237mm(Width)×133mm(Height) |
Weight | 5kg |